Function inspection – RIGOL MSO/DS2000A Series User Manual

Page 56

background image

RIGOL

MSO2000A/DS2000A Quick Guide

6

Function Inspection


1. Press StorageDefault to restore the oscilloscope to its default

configuration.

2. Connect the earth alligator clip of the probe to the “Ground Terminal” as shown

in the figure below.

3. Use the probe to connect the input terminal of CH1 and the “Compensation

Signal Output Terminal” of the oscilloscope.

Figure 7 To Use the Compensation Signal

4. Press AUTO.
5. Observe the waveform on the display. In normal condition, the display should be

a square waveform as shown in the figure below:

Figure 8 Square Waveform


6. Use the same method to test the other channels. If the square waveforms

actually shown do not match that in the figure above, please perform Probe
Compensation
.

WARNING

To avoid electric shock during the use of probe, please make sure that the
insulated wire of the probe is in good condition and do not touch the
metallic part of the probe when the probe is connected to high voltage
source.

Compensation Signal Output Terminal
Ground Terminal