beautypg.com

Input – Precision Digital PD6820 User Manual

Page 7

background image

PD6820 Loop-Powered Process Meter Instruction Manual

7

OVERALL
DIMENSIONS

5.65" x 5.25" x 4.86" (W x H x D)
(144 mm x 133 mm x 124 mm)

WEIGHT

5.00 lbs (80 oz, 2.27 kg)

WARRANTY

3 years parts and labor

Input

ACCURACY

±0.03% of calibrated span ±1 count,
square root & programmable exponent accuracy range:
10-100% of calibrated span.

ADVANCED
FUNCTION

Linear, square root, or programmable exponent

MULTI-POINT
LINEARIZATION

2 to 32 points

PROGRAMMABLE
EXPONENT

1.0001 to 2.9999

LOW FLOW CUTOFF

0-99999 (0 disables cutoff function)

TEMPERATURE
DRIFT

50 PPM/

C from -40 to 75C ambient

DECIMAL POINT

User selectable decimal point

TOTALIZER

Calculates total based on rate, time base of second, minute,
hour, or day, and field programmable multiplier; stored in non-
volatile memory upon power loss.

TOTALIZER
RESET

User selectable via SafeTouch

®

buttons, time delay, external

contact closure, or protected

CALIBRATION
RANGE

An Error message will appear if input 1 and input 2 signals
are too close together.

Input
Range

Minimum Span
Input 1 & Input 2

4-20 mA

0.10 mA

MAXIMUM
VOLTAGE DROP

Without Backlight or with
Externally-Powered (DC
Powered) Backlight

With Loop-Powered
Backlight

3.0 VDC @ 20 mA

6.0 VDC @ 20 mA

EQUIVALENT
RESISTANCE

150 Ω @ 20 mA

300 Ω @ 20 mA

EXTERNALLY
POWERED
BACKLIGHT

Voltage Range:

Maximum Power

9-36 VDC

9 VDC 12 VDC 24 VDC 36 VDC

0.2 W 0.25 W 0.5 W

0.75 W

INPUT

OVERLOAD

Over current protection to 2 A max.

This manual is related to the following products: