beautypg.com

Connections (continued) – Precision Digital PD6820 User Manual

Page 13

background image

PD6820 Loop-Powered Process Meter Instruction Manual

13

Connections (continued)

SIGNAL +

4-20 mA signal input positive terminal connection

SIGNAL -

4-20 mA signal return/negative terminal connection when not
using loop powered backlight.

BACKLIGHT +

+9-30 VDC when powering backlight from external supply.

BACKLIGHT -

4-20 mA signal return/negative terminal when using the installed
loop powered backlight or ground/negative when powering
backlight from external supply.

OUTPUT+

NPN open collector output positive.

OUTPUT-

NPN open collector output negative.

RESET +

Contact closure reset pullup to 3 VDC

RESET-

Contact closure reset ground/negative.

Refer to Figure 1 for terminal positions.

WARNING

Observe all safety regulations. Electrical wiring should be
performed in accordance with all agency requirements and
applicable national, state, and local codes to prevent damage to
the meter and ensure personnel safety.

SAFE-TOUCH

BUTTONS

BACKLIGHT

POWER

UNLOCK

LOCK

LOOP

9-36 VDC

+ - + -

SIGNAL

BACKLIGHT

+ - + -

OUTPUT

RESET

Figure 1. Connector Board