beautypg.com

Specifications, General – Precision Digital PD6820 User Manual

Page 6

background image

PD6820 Loop-Powered Process Meter Instruction Manual

6

SPECIFICATIONS

Except where noted all specifications apply to operation at +25°C.

General

DISPLAY

Five digits
(-9999 to 99999)

0.70" (17.8 mm) high, 7-segment,
automatic lead zero blanking.

Seven characters
(Total and/or Tag)

0.4" (10.2 mm) high, 14 segment.
7-digit Totalizer 9,999,999

Symbols

High, Low, & Set Alarm, Password Lock

Backlight White

DISPLAY
UPDATE RATE

Ambient > -25°C: 2 Updates/Second
Ambient < -25°C: 1 Update/5 Seconds

OVERRANGE

Display flashes

99999

UNDERRANGE

Display flashes

-9999

PROGRAMMING
METHOD

Four SafeTouch

®

through-glass buttons when cover is

installed. Four internal pushbuttons when cover is removed.

NOISE FILTER

Programmable LO, med, HI, or OFF

RECALIBRATION

Recalibration is recommended at least every 12 months.

MAX/MIN
DISPLAY

Max/Min readings reached by the process are stored until
reset by the user or until power to the meter is turned off.

PASSWORD

Programmable password restricts modification of
programmed settings.

NON-VOLATILE
MEMORY

All programmed settings and total reading are stored in non-
volatile memory for a minimum of ten years if power is lost.

NORMAL MODE
REJECTION

64 dB at 50/60 Hz

ENVIRONMENTAL

Operating temperature range: -40 to 75°C
Storage temperature range: -40 to 75°C
Relative humidity: 0 to 90% non-condensing

CONNECTIONS

Screw terminals accept 12 to 22 AWG wire

ENCLOSURE

Explosion-proof die cast aluminum with glass window,
corrosion resistant epoxy coating, color: blue. NEMA 4X, 7, &
9, IP68.
Two ¾" NPT threaded conduit openings. One ¾" NPT metal
conduit plug with 12 mm hex key fitting installed.

MOUNTING

May be mounted directly to conduit. Two slotted flanges for
wall mounting or NPS 1½" to 2½" or DN 40 to 65 mm pipe
mounting. See Mounting Dimensions on page 47.

This manual is related to the following products: