6.5 Used as a filler


Switch on PF22 and press .

Example of filling job:

Volume: 15.5 ml.
Tube: 3.0.
Velocity: 200 rpm.
Acceleration: 100.
Small/low back suction.
100 fills to be completed.
A delay of 1.1 seconds between the fills.
Product with specific gravity of 1.0
The above job is to be programmed as follows:

Volume:

<1>++<1>+<5>+<.>+<5>+

Tube:

<2>++<3>+<.>+<0>+

Velocity:

<3>++<2>+<0>+<0>+

Acceleration:

<4>++<1>+<0>+<0>+

Back suction:

<5>++<1>+

Number of fills:

<6>++<1>+<0>+<0>+

Delay:

<7>++<1>+<.>+<1>+

Specific Gravity

<9>++<1>+<.>+<0>+


You have now programmed the PF22 for the job, but want to reset the built- in counters.
In function 8 the counters indicate "number of completed fills".

Number of fills:

<8>+++


Now inform PF22 that you want to enter filling mode by pressing .
Every time or the foot switch is pressed, or every time an electric signal is given, PF22 will
run 100 fills of 8.5 ml with a delay of 1.1 seconds between the fills.

Press , verify that PF22 starts and let it run.
Press ARROW DOWN until you see function 8 on the status line and that F8: is counting each fill.

F10: gives a current indication of the number of fills completed per minute.

The fills have not been calibrated. For this function, please see chapter 5 of this manual.