5 used as a filler, Operators manual – Watson-Marlow PF22 User Manual
Page 24

Operators Manual
PF22
PF22 OM EN 74-132-005 v1.02.doc
Version: 1.02
Page 24 of 34
6.5 Used as a filler
Switch on PF22 and press
Example of filling job:
Volume: 15.5 ml.
Tube: 3.0.
Velocity: 200 rpm.
Acceleration: 100.
Small/low back suction.
100 fills to be completed.
A delay of 1.1 seconds between the fills.
Product with specific gravity of 1.0
The above job is to be programmed as follows:
Volume:
<1>+
Tube:
<2>+
Velocity:
<3>+
Acceleration:
<4>+
Back suction:
<5>+
Number of fills:
<6>+
Delay:
<7>+
Specific Gravity
<9>+
You have now programmed the PF22 for the job, but want to reset the built- in counters.
In function 8 the counters indicate "number of completed fills".
Number of fills:
<8>+
Now inform PF22 that you want to enter filling mode by pressing
Every time
run 100 fills of 8.5 ml with a delay of 1.1 seconds between the fills.
Press
Press ARROW DOWN until you see function 8 on the status line and that F8: is counting each fill.
F10: gives a current indication of the number of fills completed per minute.
The fills have not been calibrated. For this function, please see chapter 5 of this manual.