8 Interface and change of voltage


8.1 PD12 interface


1


2



3


4

1 = External 1:

PIN 1:

INPUT FOR START SIGNAL
+5 - 50 VDC, min. 100 msec. positive-edge-trigged.

PIN 2:

OUTPUT, +24 VDC, MAX. 500 MA.

PIN 3:

GROUND.

PIN 4:

STATUS OUTPUT, MAX. +24 VDC, 100 MA.
Pin 4 is grounded via an open collector during filling.

PIN 5:

STATUS OUTPUT, MAX. +24VDC, 100 mA Pin 5 is complementary to pin 4.

2 = External 2:

PIN 1:

INPUT FOR DISABLING.
+5 - 50 VDC. if this pin is activated, the drive will be disabled (no dispensing).

PIN 2:

OUTPUT, +24 VDC, MAX. 500 MA.

PIN 3:

GROUND.

PIN 4:

STATUS OUTPUT, MAX. +24 VDC, 100 MA.
Pin 4 is grounded via an open collector during filling.

PIN 5:

STATUS OUTPUT, MAX. + 24 VDC, 100 MA.
Pin 5 is complementary to pin 4.

3 = Net 1

This socket is reserved for (RS-485) network communication.

4 =Net 2

This socket is reserved for (RS-485) network communication.