5 choice of parameters, 1 programming principle, Function 1 – volume – Watson-Marlow PD12 IHS User Manual
Page 13: Function 2 - tube diameter, Function 3 - velocity, Choice of parameters, Programming principle, Description of pd12 specific functions/parameters, Instruction handbook

INSTRUCTION HANDBOOK
PD12IHS / PD12IDH
PD12IHS_PD12IDH IH EN 74-115-181 v2.00.doc
Version: 2.00
Page 13 of 19
5 Choice of parameters
5.1 Programming principle
The actual programming will be made on the MC12; however some parameters are PD12 specific.
5.1.1 Description of PD12 specific functions/parameters
The functions below affect filling and calibration.
If the value of the parameter is changed during filling or calibration, the new value will not be applied
until the start of the next filling or calibration.
Please note that if parameters are changed during filling a new calibration is reccomended.
Function 1 – Volume
Value: ml
Function 1 informs the system of the volume to be filled.
Value
Option
Min
Max
0.01
9999.9
ml. or gram
The entered value must be between 0.01 and 9999.9.
PD12 IHS:
When entering a volume of e.g. 100 ml, this will be shared among the double heads into 2 x 50 ml.
PD12 IDH:
When entering a volume of e.g. 100 ml, this will trigger each head to fill 100 ml.
Function 2 - Tube diameter
Value: Inside diameter of the tube in mm.
Drive type
Tube inside diameter in mm
PD12
0.5
0.8
1.2
1.6
3.2
4.8
6.0
8.0
Function 3 - Velocity
Value: Revolutions per minute (rpm).
Velocity range depends on tube size applied.
Range:
Tube Sizes
Max. Velocity
Max. acceleration
0.5 - 0.8 – 1.2 – 1.6
400
100
3.2
400
100
4.8 - 6.0 – 8.0
400
100