beautypg.com

Watson-Marlow 605Di User Manual

Page 10

background image

9

Tachometer output

This facility can be used to indicate motor speed or total the number of motor revolutions.

Footswitch
A footswitch or hand switch will initiate the dose. If not supplied by Watson-Marlow then select "Other" in the pumps software.

5V

0-5V DC

Care and maintenance

The only scheduled maintenance required for the 605Di is inspection of the motor brushes and their replacement before their length

is less than 10mm. The life of the brushes will depend on the duty of the pump, but is expected to be a minimum of 4,000 hours at

maximum speed.
If the pump requires cleaning, use a mild solution of detergent in water. Do not use strong solvents.
For gearbox rebuilds, use Lubriplate GR-132 (Bodine reference LG-23) only. This is a lithium combination type thickener, NL GI

No.1 grade, non-corrosive extreme pressure lubricant. This product is water resistant and resistant to a large degree to most other

contaminants.

Guard safety warnings

The 605Di will be remotely stopped and display a warning if the guard is opened during operation. This safety feature is enabled

under Manual, Dose, Analogue and RS232 operation.

Specification

Maximum rotor speed

200rpm

Voltage/frequency 100-120/220-240V

50/60Hz

Control range

50:1

Power consumption

300VA

Fuse rating

Type T (anti-surge) 5A

Operating temperature range

5 to 40C

Storage temperature range

-40C to 70C

Weight 605Di/R

24.75kg

Weight 605Di/L

30.75kg

Noise

<72dBA at 1m

Standards

IEC 335-1, EN60529 (IP55)

Machinery Directive 89/392/EEC EN60204- 1

Low Voltage Directive 73/23/EEC EN61010- 1

EMC Directive:89/336/EEC EN50081-1 / EN50082-1

Specific drive performance details such as loaded drive speed variation against mains supply voltage fluctuation and drive stability

from a cold start to normal operating temperature are available on request. For further information please contact Watson-Marlow

Technical Support.