9
Tachometer output
This facility can be used to indicate motor speed or total the number of motor revolutions.
Footswitch
A footswitch or hand switch will initiate the dose. If not supplied by Watson-Marlow then select "Other" in the pumps software.
5V
0-5V DC
Care and maintenance
The only scheduled maintenance required for the 605Di is inspection of the motor brushes and their replacement before their length
is less than 10mm. The life of the brushes will depend on the duty of the pump, but is expected to be a minimum of 4,000 hours at
maximum speed.
If the pump requires cleaning, use a mild solution of detergent in water. Do not use strong solvents.
For gearbox rebuilds, use Lubriplate GR-132 (Bodine reference LG-23) only. This is a lithium combination type thickener, NL GI
No.1 grade, non-corrosive extreme pressure lubricant. This product is water resistant and resistant to a large degree to most other
contaminants.
Guard safety warnings
The 605Di will be remotely stopped and display a warning if the guard is opened during operation. This safety feature is enabled
under Manual, Dose, Analogue and RS232 operation.
Specification
Maximum rotor speed
200rpm
Voltage/frequency 100-120/220-240V
50/60Hz
Control range
50:1
Power consumption
300VA
Fuse rating
Type T (anti-surge) 5A
Operating temperature range
5 to 40C
Storage temperature range
-40C to 70C
Weight 605Di/R
24.75kg
Weight 605Di/L
30.75kg
Noise
<72dBA at 1m
Standards
IEC 335-1, EN60529 (IP55)
Machinery Directive 89/392/EEC EN60204- 1
Low Voltage Directive 73/23/EEC EN61010- 1
EMC Directive:89/336/EEC EN50081-1 / EN50082-1
Specific drive performance details such as loaded drive speed variation against mains supply voltage fluctuation and drive stability
from a cold start to normal operating temperature are available on request. For further information please contact Watson-Marlow
Technical Support.