3. Touch the Temperature Probe head to the part whose

temperature you wish to measure. Keep the probe touching
the part under test until the reading stabilizes (about 30
seconds).

4. Read the temperature in the display. The digital reading will

indicate the proper decimal point and value.

WARNING: To avoid electric shock, be sure the thermocouple
has been removed before changing to another measurement
function

Non-Contact AC Voltage Measurements



1.

Turn meter On to any setting. The AC Voltage Detector is

located at the tip of the AC current clamp at the ‘lightning
bolt’ marks. Only the outside of the clamp, in the ‘notch’ at
the clamp separation, is sensitive to AC voltage. Clamping
the jaws around a live wire will not necessarily cause the AC

Voltage Detector to sense the ‘live’ wire

.

2. If

AC voltage is present, the red AC Voltage indicator will

light.

NOTE: The conductors in electrical cords are often twisted

and the position of the ‘hot’ wire is unknown. For best
results, hold the cord in the notch of the clamp while
running it along the length of the cord.

WARNING: Risk of Electrocution. Before use, always
test the Voltage Detector on a known live circuit to
verify proper operation