beautypg.com

Dc voltage, Range, Input impedance: all ranges are 10mω – Triplett 1101a User Manual

Page 10: Ac voltage, Input impedance: all ranges are approx. 5mω, Dc current

background image

Note:
a) The following accuracy specifications are valid at 23 degrees C,
+/- 5 degrees C, Relative Humidity less than 75%
b) The specifications are in the form
" +/- (x % of reading + dgt)" where dgt is
"Least Significant
DIGIT".

DC Voltage

Range Resolution

Accuracy

200mV 0.1mV

2000mV 1mV

20V 10mV

200V 100mV

± (0.5% rdg + 2 digits)

500V

1V

± (0.8% rdg + 2 digits)

Input Impedance: All ranges are 10MΩ

Overload Protection:

500Vrms

AC Voltage

Range Resolution

Accuracy

200V 100mV
500V 1V

± (1.2% rdg + 10 digits)

Input Impedance: All ranges are approx. 5MΩ

Frequency: 40Hz to 400Hz
Overload Protection:

200mV is 230V (AC or DC Current).

All other ranges are 500Vrms.
Display: Average Value (RMS of Sine Wave)

DC Current

Range Resolution

Accuracy

2000µA 1µA

20mA 10µA

± (1.0% rdg + 2 digits)

200mA

100µA

± (1.2% rdg + 2 digits)

10A

10mA

± (2.0% rdg + 5 digits)

Overload Protection: 0.3A / 500V fuse, 10A range is unfused.

Max. Current Input: 10A (10 seconds max, 15 minute cool down)

Burden Voltage:

300mV max

10