SuperOX
TM
Operations Manual
4. Is there a leak in the zirconia substrate? To test this property, turn off the reference air
for one minute. Measure the sensor mV as indicated by the controller or a digital
voltmeter. Turn the air back on and measure the mV again. If there is a difference
greater than 5 mV, replace the sensor.
5. Should it be necessary to remove your SuperOX
TM
Sensor from a hot furnace, do so
carefully. UNDER NO CIRCUMSTANCES should it be removed faster than 4" (101.6 mm)
per 5 minutes.
6. If your atmosphere control problem cannot be resolved, our technical support staff is
available Monday through Friday, 7:00 a.m. to 6:00 p.m. Eastern Standard Time. You may
call us at (800) 666-4330.
- Bazooka Probe (10 pages)
- Gold Probe (16 pages)
- HP2000 With 9100 Controller (10 pages)
- HP15 (23 pages)
- PGA3000 (16 pages)
- PGA3500 (26 pages)
- e-TRIM (27 pages)
- 9120 with TS (80 pages)
- MGA6000 (42 pages)
- DP2000 (17 pages)
- DPC3500 (5 pages)
- MGA6010 (54 pages)
- DPC2530 (17 pages)
- Simple Dew (18 pages)
- DPL4000 (16 pages)
- H2 Sensor (17 pages)
- Hydrogen Nitrider Analyzer (12 pages)
- PH2 (19 pages)
- AC20 Quick Start (5 pages)
- XGA Viewer (46 pages)
- AC20 RS485 Modbus (62 pages)
- AC20 (114 pages)
- 20Q Calibration Manual (9 pages)
- CAT-100 (51 pages)
- 7EK 31080 (32 pages)
- 7EK 31082 Calibration Manual (5 pages)
- 7EK 31081 (36 pages)
- 7EK 31082 (34 pages)
- 20PQ (170 pages)
- X5 Calibration Manual (1 page)
- 20Q (126 pages)
- 7SL (36 pages)
- X5 (74 pages)
- 9000 Series (15 pages)
- 3L Series (54 pages)
- 9015 Series (11 pages)
- 3 Series (92 pages)
- 9010 Series (66 pages)
- 9210 Series (65 pages)
- 9130 Series (174 pages)
- PC Configurator 2 Quick Start (15 pages)
- 9100 RPS (10 pages)
- 9125 Series (235 pages)
- 9120 RPS (11 pages)