5.Front Panel Operation
Set the N-Cycle Burst
Current Parameter
Waveform
Cycles
Load
Start Phase
Burst Period
N Cycle
Waveform
Source
Delay
Figure 5-18: The User Interface of N-Cycle Burst
(1)
When the output signal is Sine, Square, Ramp, Pulse or Arbitrary waveform, press
Burst
function button to enter the Burst mode.
(2)
Press
,
,
,
or
button to choose the waveform. Take Sine for
instance, press
to display the waveform and parameters. You can change the
parameters, please refer to "To set signals" on P11. Press
button again to
return to the Burst mode interface.
(3)
Press F4 to switch to N Cycle.
(4)
Press F1 to select Burst Period, set it to the desired value.
(5)
Press F3 to select Start Phase (if current waveform is Pulse, skip this step), define
the Start and the Stop Point in a waveform. The phase varies from -360° to +360°. For
an Arbitrary Waveform, 0° is the first waveform point.
(6)
Press F2 to select Delay, set the Time Delay between the Trigger Input and the
Start of the N-Cycle Burst. The minimum delay is a function of the specific burst
period, and should always be greater than 0s.
(7)
Press F5 to select NextPage, press it again to enter the next page.
(8)
Press F1 to select Cycles or Infinite. Set the number of Waveform Cycle in an
N-Cycle (from 1 to 50,000). If you choose Infinite, then a continuous waveform will
be generated which will not stop until a trigger event happens ( Trigger button is
pressed).
Note:
●
If needed, Burst Period will increase to cater to the specific number of cycles.
●
For an infinite-cycle Burst, External or Manual Trigger is needed to activate burst.
(9)
Press F2 to select the source. Internal means using the internal source. External
means using the Ext Trig/FSK/Burst connector in the rear panel to input the external
signal. Manual means using the external source, set the start and stop time by hand.
28