OWON AG-S series User Manual
Page 26

5.Front Panel Operation
Current Parameter
Carrier
Frequency
Carrier
Amplitude
Carrier Waveform
Carrier Waveform
Modulating Waveform
Load
Mod Depth
AM Frequency
Mod Type
Mod Shape
Source
Figure 5-12: The User Interface of AM
How to set the parameters of AM
(1)
Press Mod function button to enter the Modulation mode.
(2)
Press F1 to switch Mod Type to AM. If the Carrier Waveform is not Sine, the
system will switch it to Sine automatically.
(3)
Press
button to display the waveform and parameters of the Carrier
Waveform. You can change the parameters, please refer to "To Output Sine Signals"
on P11. Press
button again to return to the Modulation mode interface.
(4)
Press F5 to select the source. If the source is External, use the Modulation In
connector in the rear panel to input the external signal, the setting of AM is finished.
If you choose Internal, continue to the steps below.
(5)
Press F2 to choose Mod Shape, you can choose Sine, Square or Ramp.
(6)
Press F3 to set AM Frequency. The range is 2 mHz~20 KHz (Internal source
only).
(7)
Press F4 to set Mod Depth. The range is 0%~100%.
Term Explanation
AM Frequency:
The frequency of modulating waveform.
Mod Depth:
The Amplitude Range of modulating waveform. In the 0% Modulation, the output
amplitude is the half of the set one. In the 100% Modulation, the output amplitude
is the same with the set one. For an external source, the depth of AM is controlled by
the voltage level of the signal connected to the Modulation In connector in the rear
panel. +5V corresponds to the currently set depth 100%.