How to set the parameters of sweep – OWON AG (-F) Series User Manual
Page 35

5.Front Panel Operation
30
Current Parameter
Sweep Waveform
Load
Sweep Type
Sweep Time
Sweep Waveform
Source
Sweep Waveform
Amplitude
Sweep Waveform
Offset
Start
Frequency
Stop
Frequency
Figure 5-21: The User Interface of Sweep Mode
How to set the parameters of Sweep
(1)
When the output signal is Sine, Square or Ramp waveform, press the Mod button,
then press F2 to select Sweep to enter the Sweep mode.
(2)
Press
,
or
button to choose the waveforms. Take Sine for
instance, press
to display the waveform and parameters. You can change the
parameters, please refer to To set signals on page 12. Press
button again to
return to the Sweep mode interface.
(3)
Press F1 to set Sweep Time, the Time Span of the Sweep for which the
Frequency changes from the Start Frequency to Stop Frequency.
(4)
Press F2 to select the Sweep Type. Linear means setting the Sweep with linear
spacing; Log means setting the Sweep with logarithmic spacing.
(5)
Use Start Freq and Stop Freq or Center Freq and Freq Span to set the range of the
frequency. Press F3 to select Sta_Freq or Cen_Freq, and set the desired value.
(6)
Press F4 to select StopFreq or FreqSpan, and set the desired value.
(7)
Press F5 to select NextPage, press it again to enter the next page.
(8)
Press F1 to select the source. Internal means using the internal source. External
means using the Ext Trig/Burst/Fsk In connector in the rear panel to input the
external signal. Manual means choosing manual trigger; in Sweep interface, press the
knob on the front panel to trigger a Sweep.