2 cabling length, 3 signal line direction – Measurement Computing Isolator488 User Manual

Page 20

background image

Section 2

Getting Started

2.11

2.6.2 Cabling Length

The IEEE 488 specification allows for cabling distance between bus

devices of 2 meters (approx. 7 feet). The total worst case distance from the
two furthest devices is not allowed to exceed 20 meters (approx. 70 feet).
The addition of the Isolator488 does not increase this distance.

2.6.3 Signal Line Direction

The signal lines and their direction include....

Data Lines:

DIO1 - DIO7

Bidirectional

Handshake Lines:

DAV

Bidirectional

NRFD

Bidirectional

NDAC

Bidirectional

Command Lines:

SRQ

To Controller Bus

ATN

To Instrument Bus

EOI

Bidirectional

REN

To Instrument Bus

IFC

To Instrument Bus

The majority of the command lines have their signal direction fixed.

This forces bus control from the CONTROLLER bus.

When Attention (ATN) is asserted, the data lines, talker handshake

line (DAV) and the EOI line direction is forced from the CONTROLLER
to the INSTRUMENT port. The listener handshake lines (NRFD and
NDAC) are forced from the INSTRUMENT to the CONTROLLER port.
The commands are sent through the Isolator488 to the INSTRUMENT
port.