Specifications, Analog input, Accuracy – Measurement Computing USB-1608FS-Plus User Manual
Page 17: Analog input dc voltage measurement accuracy
![background image](https://www.manualsdir.com/files/797550/content/doc017.png)
17
Chapter 4
Specifications
All specifications are subject to change without notice.
Typical for 25°C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. General analog input specifications
Parameter
Condition
Specification
A/D converter type
16-bit successive approximation type
Number of channels
8 single-ended
Input configuration
Individual A/D per channel
Sampling method
Simultaneous
Absolute maximum input voltage
CHx IN relative to GND
±15 V max
Input impedance
100 MΩ min
Input ranges
Software-selectable per channel
±10 V, ±5 V, ±2 V, ±1 V
Sampling rate
Hardware paced
0.01 S/s to 100 kS/s, software-selectable
Throughput
Software paced
500 S/s all channels
Hardware paced (Note 1)
(400 kS/s) / (# of channels) max,
100 kS/s max for any channel
BURSTIO ≤ 32,768 total samples
(uses onboard FIFO)
(800 kS/s) / (# of channels) max,
100 kS/s max for any channel
Gain queue
Software configurable. Up to eight
elements – one gain element per unique,
ordered channel.
Resolution
16 bits
No missing codes
16 bits
Crosstalk
Signal DC to 25 kHz
–80 dB
Trigger source
Software-selectable
External digital: TRIG_IN
Note 1:
Maximum throughput when scanning in hardware paced mode is machine dependent.
Accuracy
Analog input DC voltage measurement accuracy
Table 2. Calibrated absolute accuracy
Range
Accuracy (mV)
±10 V
5.66
±5 V
2.98
±2 V
1.31
±1 V
0.68