Specifications

21

Power

Table 12. Power specifications

Parameter

Condition

Specification

Operating modes

Bus-powered, USB 5 V supply

Supply current
(see Note 1)

Suspend mode

<2.5 mA

Enumeration

<100 mA

Run mode

<500 mA

Power consumption
excluding analog and
digital outputs

Run mode

1.175 W max (235 mA input current)

Power available for +5 V,
AICKO, AOCKO, TMR,
analog outputs, and
digital I/O

Run mode

1.325 W max
The total power consumption for all external loads must be less than this
value, and each load must meet the individual specification for the
terminal.

Digital output power
calculation

Power per output = Iout × 5 V
(for example, @ 24 mA, P = 0.024 × 5 = 120 mW / output)

Analog output power
calculation

Power per output = (Iout × 16.5 V) / 0.78
(for example, @ 3 mA, P = (0.003 × 16.5) / 0.80 = 63.5 mW/output)

+5 V output power
calculation

Power (W) = Iout × 5 V

+5 V output voltage
range (see Note 2)

Run mode

4.25 V min, 5.25 V max

Suspend mode,
enumeration

0 V

+5 V output current

Run mode, no
other output loads

265 mA max (1.325 W)

Fuses

On USB supply

0452.750 - Littelfuse 0.750A NANO2® Slo-Blo®

Subminiature Surface

Mount Fuse.
Spare fuse mounted in holder on PCB.

Note 1:

This is the total current consumption for the USB-1208HS-4AO, including +5 V, digital output and
analog output currents (if supported).

Note 2:

Output voltage range assumes input power is within specified limits.

USB specifications

Table 13. USB specifications

Parameter

Specification

USB device type

USB 2.0 (high-speed)

USB device compatibility

USB 1.1, 2.0

USB cable length

5 m (16.40 ft) max

USB cable type

A-B cable, UL type AWM 2725 or equivalent (min 24 AWG VBUS/GND,
min 28 AWG D+/D–).

Environmental

Table 14. Environmental specifications

Parameter

Specification

Operating temperature
range

0 °C to 50 °C

Storage temperature
range

–40 °C to 85 °C

Humidity

0% to 90% non-condensing

This manual is related to the following products: