beautypg.com

Measurement Computing USB-1208HS-4AO User Manual

Page 17

background image

USB-1208HS-4AO User's Guide

Specifications

17

Table 2. Calibrated absolute accuracy

Range

Accuracy (mV)

±20 V (DIFF mode)

±9.55 typ, ±13.18 max

±10 V (DIFF mode)

±4.59 typ, ±6.23 max

±5 V (DIFF mode)

±2.25 typ, ±2.75 max

±10 V (SE mode)

±5.10 typ, ±8.06 max

±5 V (SE mode)

±2.63 typ, ±4.03 max

±2.5 V (SE mode)

±1.59 typ, ±2.70 max

0 – 10 V (SE mode)

±3.29 typ, ±5.13 max

Table 3 summarizes the noise performance for the USB-1208HS-4AO. Noise distribution is determined by
gathering 50 kS with inputs tied to ground at the user connector. Samples are gathered at the maximum
specified sample rate of 1 MS/s.

Table 3. Noise performance

Range

Typical counts

LSBrms

±20 V (DIFF mode)

3

0.45

±10 V (DIFF mode)

3

0.45

±5 V (DIFF mode)

3

0.45

±10 V (SE mode)

5

0.91

±5 V (SE mode)

5

0.91

±2.5 V (SE mode)

5

0.91

0 – 10 V (SE mode)

5

0.91

Table 4. Input settling time in µs, typical

Condition

Range

±1 LSB

±4 LSB

±8 LSB

+ full-scale to –full-scale
channel switch, same range
to same range

±10 V

1.5

1.1

1.0

±5 V

2.1

1.1

1.0

±2.5 V

2.2

1.1

1.0

0 V to 10 V

2.6

1.1

1.0

This manual is related to the following products: