beautypg.com

Measurement Computing CIO-EXP32 User Manual

Page 21

background image

CIO-EXP32 User's Guide

Functional Details

Output voltage vs gain at maximum temperature

Type

Min Temp °C

Max Temp °C

mV at

Max

Gain / Output Volts @ Max Temp

100

200

300

500

800

J

0

750

42.28 4.2 8.5 12.7 21.1 33.8

K

−200

1250

50.63 5.1 10.1 15.2 25.3 40.5

T

−200

350

17.82

1.8 3.6 5.3 8.9 14.3

E

−200

900

68.78 6.9 13.8 20.6 34.4 55.0

S 0

1450

14.97

1.5 3.0 4.5 7.5 12.0

R 0

1450

16.74

1.7 3.3 5.0 8.4 13.4

The table below lists the maximum readable temperatures for each gain and thermocouple type for an A/D
range of ±5V.

Gain vs maximum readable temperatures

Type

Gain / Max. Readable Temps. with A/D Range = ±5 V

10

100

200

300

500

800

J

MAX

MAX

457 °C

307 °C

186 °C

117 °C

K

MAX

1232 °C

602 °C

406 °C

246 °C

152 °C

T

MAX

MAX

MAX

330 °C

214 °C

140 °C

E

MAX

661°C

350 °C

244 °C

152 °C

98 °C

S

MAX MAX MAX MAX 1035

°C

697

°C

R

MAX MAX MAX 1445

°C

961°C 658

°C

To read the maximum usable temperature for type K and E thermocouples, set the gain to

X10

or less.

After you change the gain setting, calibrate the CIO-EXP32 using the InstaCal utility.