Specifications, Power consumption, Analog input – Measurement Computing CIO-DAS-TEMP User Manual
Page 17: Counters, Environmental, Sensors, 7 specifications
![background image](https://www.manualsdir.com/files/797414/content/doc017.png)
7 SPECIFICATIONS
POWER CONSUMPTION
+5V Normal Operation
890 mA Typical
960 mA Max
ANALOG INPUT
Channels
32 Current Input, AD592 Sensor Specific
Range
0 to 500
µΑ
Input Impedance
20 KOhm (nominal)
Resolution
30 S/S
0.015 Deg C (60 Hz Line Frequency)
25 S/S
0.0125 Deg C (50 Hz Line Frequency)
200 S/S
0.1 Deg C (400 Hz Line Frequency)
Current/Temp
1.0 Degrees Kelvin /
µAmp
Accuracy
± 0.02%
Linearity Error
± 0.02%
Gain Drift
±75 ppm/°C
V/F Converter
AD652
Calibration
0 to 5V (pin 37) @ 10,000 KOhm Input Impedance
COUNTERS
Counter Type
82C54 Three 16-Bit Down-counters, 10 MHz Max Input
Prescaler
2 MHz divided by 14 programmable steps
Interrupts
2 through 7 Programmable
ENVIRONMENTAL
Operating Range
0 to 60°C
Storage Range
−40 to 100°C
Humidity
0 to 90% non-condensing
SENSORS
AD592 Sensor
CN Suffix
BN Suffix
Temp Range
−25 to 105°C
−25 to 105°C
Accuracy at range
0 to 70°C, ± 0.8°C
0 to 70°C, ± 1.5°C
−25 to 105 °C, ± 1.0°C
−25 to 105°C, ± 2.0°C
Temperature drift
±25 ppm/°C
±25 ppm/°C
Storage Range
−45 to 125C
−45 to 125°C
13