What if, 1 error messages – Ivoclar Vivadent P300 User Manual
Page 27
27
8. What if …
This chapter will help you to recognize malfunctions and take
appropriate measures or, if possible and acceptable, to perform
some simple repairs.
8.1 Error messages
The furnace continuously checks
all functions during operation.
If an error is detected, the
respective error message is
displayed.
In case of an error, the heater switches off for safety reasons.
The following error messages may be displayed. If there are any questions, please contact the Ivoclar Vivadent After Sales Service.
Error
Conti-
Error
Error Message Text
Hint
nuation
No.
possible
2
T < B
Enter a logical value for T
8
L > T
Enter a logical value for long-term cooling L
9
V2x <= V1x
Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2
10
V2x > Tx + 1°C
Change either the vacuum values or the holding time T
11
Incorrect values for V1x, V2x
Enter a logical value for V1x, V2x
13
Current temperature after
Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down.
*, **
Start > Tx + 80 °C
14
Temperature in the firing chamber
Temperature too high for calibration; the furnace cools down! Start the program again later!
*
> 410 °C at the start of the calibration
program, i.e. too high
16
T2 < T1
Enter a lower value for T1 or a higher value for T2.
17
Power failure > 10 s during a firing
A firing program in progress was interrupted for more than 10 s. The program cannot be continued!
program in progress
18
T1 > V12
Enter a lower value for T1 or a higher value for V12
20
no
Error in the heating system
Check the heater fuse. If the fuse is i.o. contact your Service Centre.
**
23
Heating muffle very old
The heating muffle is very old. It is recommended to replace it. After the error message has been
acknowledged, a firing program may still be started.
24
Heating muffle defective
The condition of the muffle is so poor that is has to be replaced immediately.
26
T is > B + 160 °C at the start of a
Firing chamber too hot to start a firing program.
firing program
27
no
Furnace head cannot
The furnace head cannot be moved to the final position. It might be blocked by an external mechanical
**, ***
be initialized
source! If not, contact your Service Centre.
28
The furnace head does not
The furnace head does not open/close correctly. The furnace head was manually moved or is obstructed.
**
reach the target position
The furnace head must only be moved using the keys intended for this purpose!
32
no
The vacuum is not released
The vacuum cannot be released. The vacuum valve might be dirty or stuck. Contact your Service Center.
**
33
Necessary vacuum (xxxmbar) is not
The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum pump,
reached within 1 min.
pump fuse.
110
HV > H (H2)
Enter a lower value for HV or a higher value for H (H2)
702
Brief power failure during a program
A program in progress was interrupted by a brief power failure. The program is continued!
in progress
800
Final vacuum value not reached
The required final vacuum value cannot be reached. Check the vacuum pump.
801
Vacuum drop
An unacceptable vacuum drop has occurred.
802
The vacuum does not increase
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
(self-test)
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
1510
Temperature > VT at the start of a
The temperature in the firing chamber is higher than the pre-drying temperature. Press START to continue
firing program
the program despite the error message.
1700
Service note
Two years have passed or the heating muffle has been in use for more than 1200 firing hours since the last
inspection of the furnace. Ivoclar Vivadent therefore recommends that your furnace should be inspected.
Please refer to the Equipment Service Passport or the Operating Instructions for further information. The
interval until the next appearance of the note can be set in the Settings.
*
Furnace head opens when this error occurs.
**
A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.