7. Maintenance, Cleaning and Diagnosis
This chapter describes the user maintenance and cleaning procedures
for the Programat P300/G2. All the other tasks must be performed
by qualified service personnel at a certified Ivoclar Vivadent Service
Centre.
7.1 Monitoring and maintenance
The time for these maintenance procedures depends on the
frequency of use and the working habits of the users. For that
reason, the recommended times are only approximations.
monthly
Vacuum hose
Firing chamber
Check if there is condensate in the vacuum hose or firing chamber.
daily
Firing chamber
Check the quartz glass cylinder to make sure the quartz glass is not defective.
twice a
year
Firing chamber
Check temperature.
Use the temperature checking set to check and adjust the temperature in the
furnace.
weekly
Keypad (10)
Check the keypad for visible damage. If the keypad is damaged, it has to be replaced
by a certified Ivoclar Vivadent Service Centre.
weekly
Sealing rims of the furnace
head (2) and the furnace base
(1)
Check if the sealing rims of the furnace head and the furnace base are clean and
undamaged.
monthly
Insulation (3)
Check the insulation for cracks and damages. If the insulation is worn down it has to
be replaced by a certified Ivoclar Vivadent Service Center. Fine hairline cracks on the
surface of the insulation are harmless and do not influence the function of the
furnace in a negative fashion.
weekly
monthly
weekly
When
Thermocouple (4)
Check if the thermocouple is straight and in the right place.
Opening mechanism
Check if the furnace head opens smoothly and without excessive noise.
Var. external connections
Check all plug-in connections for correct fit
What
Part
This furnace has been developed for typical use in dental
laboratories. If the product is used in a production
enterprise, for industrial applications and for continuous
use, premature ageing of the expendable parts has to be
expected.
The expendable parts are as follows:
– Heating muffle
– Insulation material
Expendable parts are not covered by the warranty. Please also
observe the shorter service and maintenance intervals.
In general, the furnace head should not be replaced since the components
(furnace head and furnace base) have been coordinated with each other.
However, if the furnace head must be replaced for maintenance reasons,
subsequent temperature calibration is required.
7.2 Cleaning
The furnace may only be cleaned when it is cool, since
there is a burn hazard. Do not use any cleaning solutions.
The following parts have to be cleaned from time to time:
*Never clean with compressed air!
Item:
Frequency:
Cleaning material:
Housing (9) and furnace
if required
soft, dry cloth
head (25)
Keypad (10)
weekly
soft, dry cloth
Cooling tray
daily
cleaning brush *
(34)
Insulation (3)
daily
cleaning brush *
Sealing rim of the furnace daily
cleaning brush and
head (2) and sealing
a soft cloth
surface (1)