p3

Instructions

The following detail the steps for probing a
membrane using the ImmunoBlot.

Select and block membrane

1

Select a membrane of choice (nitrocellulose,
Hybond ECL

; PVDF, Hybond P; low fluorescent PVDF,

Hybond LFP) for the assay and cut to:
Length: 15 cm
Width: Cut to accommodate channels to be used

Maximum width for all channels: 15 cm

2

Block membranes with excess non-specific protein.

1

3

Note: If blocked membranes have been stored in a
protein-free solution, re-block them for 1 to 2 minutes
in a tray of blocking solution.

Load membrane

1

Remove the top acrylic plate of the ImmunoBlot
and turn it over.

2

With the antigen-bearing face of the membrane facing
the ImmunoBlot channels, position the membrane so
it covers all the channels.

2

3

Place a new, dry sealing pad (supplied with unit) over
the membrane.

4

Set the bottom plate of the unit on the inverted top
plate so the alignment pins fall in place.

5

Ensure there is no gap between the top and
bottom plates.

1

See Appendix A, Note 1,
Blocking the membrane.

2

See Appendix A, Note 2,
Aligning the membrane.