beautypg.com

Troubleshooting – Hoefer SE250 User Manual

Page 23

background image

5. Troubleshooting

problem

solution

Smile effect on the buffer front

To reduce the running temperature:

• Circulate coolant through the upper buffer

chamber core.

• Prechill the buffer.

• Decrease the current or voltage setting. (10 mA per

0.75 mm gel, 15 mA per 1.5 mm thick gel.)

• Run the gel in the cold room.

Protein streaks vertically

• Centrifuge or filter sample before loading to remove

particulates.

• Dialyze or desalt the sample.

Unusually slow (or fast) run

Adjust the solutions:

• Check recipes, gel concentrations, solutions, and

dilutions. (For instance, do not use Tris-HCl instead
of Tris.)

• If the required pH of a solution is exceeded, do not

back-titrate. Prepare fresh buffer.

• Dispose of older acrylamide solutions and use only

stock of the highest quality.

• Only use freshly deionized urea.

Adjust the voltage or current settings:

• To increase or decrease the migration rate, adjust

the voltage or current by 25–50%.

Bands are skewed or distorted

Check gel preparation and polymerization:

• Degas the stacking gel solution and avoid trapping

air bubbles under the comb teeth.

• Overlay the running gel with water-saturated

n-butanol before polymerization begins to avoid
forming an uneven gel surface.

Check sample preparation:

• Dialyze or desalt the sample.

• Centrifuge or filter sample before loading to remove

particulates.

p15

This manual is related to the following products: