beautypg.com

Hoefer SE250 User Manual

Page 21

background image

After the run

1

Once the tracking dye reaches the bottom of the gel,
turn off the power supply, disconnect the leads, and
remove the safety lid.

2

If coolant is circulating, stop the flow and disconnect
the fittings or tubing.

3

Remove the core assembly with gels attached by
squeezing the release tabs and lift out the core
assembly.

4

Pour out the buffer by inverting the core assembly,
then remove both clamps, and lift away gel
sandwich(es) from the upper buffer chamber core.

5

Gently loosen and then slide away both spacers. Slip an
extra spacer or a Hoefer Wonder Wedge into the bottom
edge (to prevent breaking the ears of the notched
plates) and separate the plates. The gel usually adheres
to the alumina plate. Carefully lift the gel from the plate
and lay it into a tray of stain or fixative.

Important!  Always disconnect
the high voltage leads from the
power supply before removing
the lid from the unit.

p13