CMG-3T

Most data can be found on the calibration sheet for your sensor.
Under Instrument response, you should fill in the sensor

response code for your sensor, according to the table below.
Instrument Type should be set to the model number of the

sensor.

If the file calvals.txt exists in the same directory as
Scream!'s executable (scream.exe), Scream! will look there for

suitable calibration values. A sample calvals.txt is supplied
with Scream!, which you can edit to your requirements. Each

stream has its own section in the file, headed by the line
[instrument-id]. The instrument-id is the string which

identifies the digitizer in the left-hand pane, e.g. GURALP-DEMO.
It is always 6 characters (the system identifier) followed by a

dash, then 4 characters (the serial number.) For example:

[instrument-id]

Serial-Nos=T3X99

VPC=3.153,3.147,3.159

G=1010,1007,1002

COILCONST=0.02575,0.01778,0.01774

CALVPC=3.161

CALRES=51000

TYPE=sensor-type

RESPONSE=response-code

9. Click OK. The script will return with a graph showing the

responsivity of the sensor in terms of amplitude and phase plots
for each component (if appropriate.)

The accuracy of the results depends on the amount of data you

have selected, and its sample rate. To obtain good-quality
results at low frequency, it will save computation time to use

data collected at a lower sample rate; although the same

26