12 some of the most important photon energies, Cd109, Bi214 – GBS Elektronik MCA 166 User Manual
Page 45: Ce139, Co60, Co57, Cs137, Eu152, Hg203, Mn54
_______________________________________________________________________
45
3.11 Some examples for parameters of MCA used
with different detectors
This table shall just give an overview of possible settings useful with the MCA. It is far
from being complete. For actual settings with your detector, see the detectors manual.
Type
Model
Manufacturer
HV (V)
+
12
V
-
12
V
+
24
V
-
24
V
Input
signal
polarity
Channels
Gain for 2
MeV
full
range
Shap-
ing
time
(µs)
CdZnTe
SDP/Z/60
Ritec
+300
x
x
-
-
neg
1024
90
1
CdZnTe
SDP/Z/20
Ritec
+500
x
x
-
-
neg
1024
80
1
CdZnTe
CZT500
Ritec
+600
x
x
-
-
pos
1024
128
1
NaI
Scintiblock
Crismatec
+500
pos
512
7.2
1
NaI
25825
Scionix
+550
x
x
-
-
pos
512
90
1
NaI
72016
72000
Robotron
Messelektroni
k
-800
neg
512
14
1
CsI
Sc1010
Eurorad
-
x
x
-
-
neg
256
160
1
HPGe
GMX-20190-
S
Ortec
-3000
neg
4096
18
2
HPGe
GEM 15-190
Ortec
+3000
pos
4096
24
2
HPGe
GR 2018
Canberra
-3000
neg
4096
11
2
HPGe
GL 0310
Canberra
-2000
x
x
x
-
neg
4096
2.3
2
Si
planar
7865S
Ortec
-1000
neg
4096
1.9
2
HPGe
EGC 30-190R Eurisys
-3000
pos
4096
52
2
HPGe
EGPC 25-185 Eurisys
+2000
neg
4096
35
2
3.12 Some of the most important photon energies
Isotope
Half-life
Energy (keV)
Branching ratio (%)
Am241
432.2y
26.34/59.54
2.4/36
Cd109
462.6d
88.03
3.63
Bi214
19.9min, daughter
of Ra226
609.31
46.09
Ce139
137.64d
165.86
79.87
Co60
5.271y
1173.23/1332.49
99.85/99.98
Co57
271.8d
14.4/122.06/136.47
9.16/85.6/10.68
Cs137
30.07y
661.66
85.1
Eu152
13.3.y
121.77/344.28/778.91/
964.11/1112.07/1408.0
0
28.38/26.59/12.98/14.46/1
3.57/20.85
Hg203
46.61d
279.19
81.84
Mn54
312.2d
834.82
99.98
Pb214
26.8min, daughter
of Ra 226
241.91/295.09/351.86
7.46/19.17/37.06
Ra226
1600y
186.11
3.28
Sn113
115.09d
391.7
64.89
Sr85
64.84d
514.0
98.4
U235
7.04E8y
143.78/163.37/185.73/
205.33
10.53/4.7/53.15/4.7
Y88
106.65
898.04
94