Fluke Biomedical DALE601 User Manual
Page 51

Maintenance and Service
Verifying Analyzer Performance
3
3-5
Table 3-1 Performance Verification Parameters (cont.)
Parameter Procedure
Desired
Reading
M.A.P. (lead
isolation current)
Connect the black chassis
cable with a clamp with
black insulation to the
EXTERNAL connector and
the clip to the lead
selected. Press ISO
TEST.
1000 ±15 %, depending on
line voltage
External; for
leakage current
measurement
Connect the black external
cable with a clamp with
red insulation to the
EXTERNAL connector and
the clip to the 100 µA test
point.
100 ±2 µA
Isolation
Clip the black external
cable with a clamp with
red insulation to the black
chassis cable and press
the ISO TEST switch.
1000 ±15 %, depending on
the line voltage
Calibration
Do not attempt.
Not Applicable
Table 3-2. Current Verification Parameters
Current (Amps)
Bulb (Watts)
120 V Line
230 V line
60 0.50 0.30
75 0.63 0.35
100 0.83 0.45
150 1.25 0.70