Preparing the analyzer for use – Fluke Biomedical DALE601 User Manual
Page 28
DALE601/601E
Operators Manual
2-4
Preparing the Analyzer for Use
XW
Warning
To avoid possible electrical shock or personal injury to the
patient, disconnect all patient connections to the device to
be tested before starting.
To prepare the Analyzer for use:
1. Place the Analyzer’s switches in the following initial positions:
FUNCTION switch
Mains Voltage
NEUTRAL switch
CLOSED
OUTLET switch
OFF (center)
2. Plug the Analyzer into a properly-rated outlet.
XW
Warning
To avoid possible electrical shock or personal injury to the
patient, plug the Analyzer only into a grounded, “Hospital
Grade” power receptacle. Do not circumvent this
precaution for any reason.
The Analyzer is equipped with a hospital-grade power plug. Grounding
reliability can be achieved only when the Analyzer is connected to an
equivalent power receptacle marked "Hospital Grade." Grounding is
important for personnel safety and to make some of the tests offered by
the Analyzer.
Note
This device is sensitive to electromagnetic fields. If interference is
encountered in measurements, reconfigure the test setup or contact
the manufacturer.