Fluke Biomedical DALE601 User Manual
Page 33
Operation
Measuring Earth Resistance
2
2-9
1. Set the
OUTLET
switch to the
OFF
position.
2. Connect the standard cable (supplied with unit) that has a black coil cord
cable with the clamp with black insulation to the
CHASSIS
connector on
the top panel of the Analyzer.
3. Clamp the cable clip to the exposed chassis of the DUT or to the
enclosure, if conductive.
Note
Ensure that bare metal is reached and that both jaws of the clip are in
contact with the chassis. Do not use metal labels or incidental
conductive hardware for this test.
4. Once connection is made, set the
FUNCTION
switch to
EARTH
RESISTANCE
.
MAI
N
S
V
OLTAGE
CURRE
N
T AMPS
EARTH RESISTA
N
CE
EARTH LEAKAGE
E
N
CLOSURE LEAKAGE
PATIE
N
T LEAD LEAKAGE
PATIE
N
T AUX CURRE
N
T
M.A.P.
EXTER
N
AL
fcu05.eps
The earth resistance value is displayed to 19.99
Ω
.