Resistance, Xw warning – Fluke Biomedical DALE601 User Manual
Page 45
Operation
External
Measurements
2
2-21
2. Set the
FUNCTION
switch to
EXTERNAL
.
MAI
N
S
V
OLTAGE
CURRE
N
T AMPS
EARTH RESISTA
N
CE
EARTH LEAKAGE
E
N
CLOSURE LEAKAGE
PATIE
N
T LEAD LEAKAGE
PATIE
N
T AUX CURRE
N
T
M.A.P.
EXTER
N
AL
fcu11.eps
3. Press the
ISO TEST
switch.
This action applies isolated line voltage between the two sides. The
Analyzer displays the isolation current that flows.
Note
The method for making connection to either side of the isolation
barrier varies with the device to be tested.
XW
Warning
To avoid possible electrical shock or personal injury
take
care when handling the cables. High voltage, 110 percent
of line volts, will be accessible between the two cables
when the LIFT GROUND/ISO TEST switch is in the ISO
TEST position.
Resistance
Resistance measurements between two points are made to verify the integrity
of permanently installed equipment whose ground cannot be broken to
measure the chassis leakage current. These are usually high power devices,
which can have high leakage currents and depend on the bonding of all chassis
to a common point for safety. It is recommended that these devices be powered
off for testing.