Fluke Biomedical DALE601 User Manual
Page 21

Introduction and Specifications
Specifications
1
1-11
MAP / Ground Switch ......................................... 2 position rocker (Open Ground /
Apply Mains on Applied Part test
voltage)
DUT Power............................................................. 15 A @ 120 V, 20 A @ 120 V,
10 A @ 230 V
Physical Size ......................................................... 4.0 x 7.8 x 1.6 in
(102 x 196 x 41 mm)
Mains Voltage
Range ................................................................. Line Voltage
Accuracy ............................................................. ±2 % of reading, ±1 digit
Current Consumption
Range ................................................................. 0 – 14.99 A (Dale601-1515)
0 – 19.99 A (Dale601-1520/2020)
0 – 9.99 A (Dale601E)
Accuracy ............................................................. ±5 % of reading, ±1 digit
Earth Resistance
Range ................................................................. 0 – 19.99
Ω
Accuracy ............................................................. ±1 % of reading, ±1 digit
(0 – 1.99
Ω
)
±2 % of reading, ±1 digit
(2 – 19.99
Ω
)
Current Source...................................................... 10 mA DC (four-wire bridge)
Leakage Current
Range ................................................................. 0 – 1999 µA
Accuracy
DC and 25 HZ to 1 KHz ................................. ±1 % of reading, ±3 LSD
1.0 KHz to 100 KHz ........................................ ±2.5 % of reading, ±3 LSD
100 KHz to 1 MHz .......................................... ±5 % of reading, ±3 LSD
Type Measurement......................................... True rms
Test Load........................................................ AAMI or IEC601 – 1 (selectable)
Isolation Test (Mains on Applied Parts)
Voltage................................................................ Mains voltage
Current................................................................ Limited @ 1 mA
Current Capacity
DALE601-1515 ................................................... 15 A
DALE601E .......................................................... 10 A
DALE601-1520/2020 .......................................... 20 A, 20 % duty cycle
(2 minutes on, 8 minutes off)
Environmental
Operating Temperature ...................................... 15 – 40 ºC
Storage Temperature.......................................... - 20 – 65 ºC
Relative Humidity................................................ 90 % max