Kvp measurement procedure, Selecting a filter pack, Using the x-ray kvp function – Fluke Biomedical 199XRAY User Manual
Page 9: Chapter 2

2-1
Chapter 2
Using the X-RAY kVp Function
kVp Measurement Procedure
To perform kVp measurements, do the following:
1
Follow the operating instructions for the filter
pack as described in the filter pack Operators
Manual.
Connect the kVp divider to Input A of the
199XRAY using a shielded 1:1 cable, for
example the cable of the MA190 Accessory set.
2
Select the required filter pack, see ‘Selecting a
Filter Pack’.
3
Start the measurement. See kVp Scope Mode
on page 2-7 for further instructions.
Selecting a Filter Pack
To select a filter pack, do the following:
1
Display the
MEDICAL/RECORDER screen.
2
Select ■ kVp with the arrow keys,
then accept with the F4 (ENTER)
key. One of the Filter Pack
Selection Screens will be shown.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)