Kvp scope mode screen, readings, f1-f4 keys, Kvp mode readings – Fluke Biomedical 199XRAY User Manual
Page 18

Fluke 199XRAY 
Users Manual 
2-10
kVp Scope Mode Screen, Readings, F1-F4 Keys
An example of the kVp scope mode screen is shown in the 
figure below. Numerical readings and the softkey button 
bar differ from the normal Scope mode. 
kVp Mode Readings
Voltage readings:
xxx.x kVp - avg
average kVp value between cursors,
indicated by the tick marks on the 
vertical cursors 
xxx.x kV - peak
kV peak value between cursors
In the MANUAL mode other readings may be shown 
depending on the selected cursors. See ‘F1…F4 softkey 
functions for cursor menu ON’ on the next page. 
TIME/PULSE# Readings:
You can select Time or Pulse readings, if applicable, using 
the F3 key. 
xx.xx
ms
time of pulse (AUTO)
xx.xx
ms
time of multiple pulses (AUTO)
xxx
pls
number of pulses (AUTO)
xx.xx
ms
time between cursors (MANUAL)
xxx
pls
number of pulses between cursors 
(MANUAL) 
Note: If the kV or exposure time values cannot be
calculated because of the signal shape, or because 
the kV is not within the range of the chosen filter 
pack, dashes will be displayed. Use the Manual 
mode to read the voltage at each cursor position or 
adjust the vertical and/or horizontal sensitivity to 
obtain a signal that meets the Auto kVp Mode 
Waveform Requirements. 