Kvp scope mode presets – Fluke Biomedical 199XRAY User Manual
Page 20

Fluke 199XRAY
Users Manual
2-12
kVp Scope Mode Presets
When entering the kVp mode some Scope functions are
preset. Some of these function setting are fixed, some can
be changed. See Table 2-1 and Table 2-2 for details.
Trigger settings can not be changed.
Vertical defection can be changed using the RANGE
key.
Time base can be changed using the TIME key. The
highest time base speed is 500 μs/div.
Sensitivity, Probe Factor, and Bandwidth filter are
fixed.
Note:
In kVp mode Input B can be used , but will show
no readings. A trace will only be shown if
triggering via Input A occurs.
Table 2-1. Default Filter Pack Settings
Filter Pack
Vertical Sensitivity
Zero Value
Horizontal Deflection
Trigger Level
Wide Range
20 kV/div
40 kV
10 ms/div
80 kV
K-Edge Mammo
500 mV/div
0 V
10 ms/div
1 V
Linear Mammo
5 kV/div
0 kV
10 ms/div
10 kV
CT
20 kV/div
60 kV
10 ms/div
100 kV
Mobile Pack Plus
20 kV/div
40 kV
10 ms/div
80 kV
Low Range
20 kV/div
26 kV
10 ms/div
46 kV
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)