Fluke Biomedical 199XRAY User Manual
Page 21

Using the X-RAY kVp Function
kVp Scope Mode
2
2-13
Table 2-2. kV Range and Trigger Level vs. Filter Pack and Vertical Sensitivity
Filter Pack
Vertical Sensitivity
kV Range
Trigger Level
Wide Range
20 kV/div *
80 -150 kV
80 kV
10 kV/div
60 – 100 kV
60 kV
5 kV/div
50 – 70 kV
50 kV
K-Edge Mammo
500 mV/div*
N/A
1 V
Linear Mammo
5 kV/div *
10 – 30 kV
10 kV
10 kV/div
20 – 40 kV
20 kV
CT
20 kV/div*
100 – 140 kV
100 kV
10 kV/div
80 -120 kV
80 kV
5 kV/div
70 – 90 kV
70 kV
Mobile Pack Plus
20 kV/div*
80 – 135 kV
80 kV
10 kV/div
60 – 100 kV
60 kV
5 kV/div
50 – 70 kV
50 kV
Low Range
10 kV/div*
46 – 90 kV
46 kV
5 kV/div
36 – 60 kV
36 kV
2 kV/div
30 – 40 kV
30 kV
* Default settings.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)