2-4
If the exposure is within the range 27.5 to 29.5 kVp, the 
reading is accurate. If ---.kVp is displayed, set the 
generator for 28 kVp and repeat step 4. If 0.0 kVp is 
displayed, set the generator to 30 kVp and repeat step 4. 
The k-edge kVp is used to determine linear filter pack 
correction. 
Once an accurate determination of kVp has been made do 
not disturb the x-ray generator settings. Proceed to linear 
mammo filter pack use. 
Linear Measurement Setup:
Remove the k-edge filter pack and replace it with the 
linear mammo filter pack. Increase the target to filter pack 
distance to about 15 inches (38cm). If the Model 
35080/A/B/M kVp Divider changes position in the x or y 
axis with respect to the x-ray generator during the linear 
kVp measurements, you must repeat the determination of 
the k-edge kVp and the correction value since this 
measurement is position sensitive. 
Linear Correction Exposure:
Make an exposure using the same generator settings as 
those used to make the last k-edge exposure. After this 
measurement is complete, the kV correction for the linear 
filter pack will be displayed. If you do not want your linear 
kVp measurements automatically corrected, press F1 
(LINEAR CHANGE…) and select No Correction. To 
recalibrate the k-edge correction, select Recalibrate… 