About the x-ray kvp function – Fluke Biomedical 199XRAY User Manual
Page 6

Fluke 199XRAY 
Users Manual 
1-2
About the X-RAY kVp Function
The 199XRAY Medical Scopemeter, when used with Fluke 
Biomedical kVp Divider, automatically measure kVp 
average and kV peak. In the radiographic measurement 
mode, exposure time is also measured in milliseconds or 
pulses. 
The Fluke Biomedical Non-Invasive kVp Divider (Model 
35080/A/B/M) is an alternative to high-voltage divider 
tanks. It provides safe, non-invasive, fast measurement of 
kVp without endangering personnel and/or equipment. Its 
use is primarily intended for X-ray machine service and 
calibration. The kVp Divider provides an analog output 
voltage proportional to the instantaneous voltage applied 
to the X-ray tube. 
The X-RAY kVp function provides oscilloscope setups for 
measurements using the kVp Divider and the following 
filter packs on Input A : 
• Wide Range Filter Pack 
• K-edge Mammo Filter Pack
• Linear Mammo Filter Pack 
• CT Filter Pack 
• Mobile Pack Plus Filter Pack
• Low Range Filter Pack
For details on the kVp Divider refer to the 35080/A/B/M 
Operators Manual. 