beautypg.com

General specifications, Detailed specifications, Oximeter spo2 optical emitter and detector – Fluke Biomedical SPOT Light User Manual

Page 26

background image

SPOT Light
Users Manual

14

General Specifications

Temperature

Operating ............................................................ 10

°

C to 40

°

C (50

°

F to 104

°

F)

Storage ............................................................... -20

°

C to +60

°

C (-4

°

F to +140

°

F)

Humidity ................................................................. 10 % to 90 % non-condensing
Altitude ................................................................... 3,000 m (9,843 ft)
Size (W x H x D) ..................................................... 12.53 cm x 14.86 cm x 4.77 cm (4.94 in x 5.85 in x 1.88 in)
Display ................................................................... LCD Monochrome display
Communication (USB Device Virtual COM Port) .. Mini B connector for service upload of firmware
Power ..................................................................... Lithium-Ion rechargeable, 3.7 V, 10.75 Wh battery, 2900 mAh
Battery Charger ..................................................... 100 V to 240 V, 50/60 Hz input, 6 V/2.5 A output. For best performance, the battery

charger must be connected to a properly grounded ac receptacle.

Battery Life ............................................................ 10 hours (minimum)
Weight .................................................................... 0.29 kg (0.7 lb)
Safety Standards ................................................... EN/IEC 61010-1:2001

Certifications ...........................................................

P, ), 

Electromagnetic Compatibility (EMC) ................. EN 61326-1:2006

Detailed Specifications

Oximeter SpO

2

Optical Emitter and Detector

%O

2

O

2

Saturations ..................................................... 80 %, 85 %, 90 %, 95 %, 97 %, 98 %, 99 %, and 100 %

Accuracy

With oximeter manufacturer’s R-curve

Saturation within UUT specific range ..........

±

(1 count + specified accuracy of the UUT)

Saturation outside UUT specific range ....... monotonic with unspecified accuracy