General specifications, Detailed specifications, Oximeter spo2 optical emitter and detector – Fluke Biomedical SPOT Light User Manual
Page 26
SPOT Light
Users Manual
14
General Specifications
Temperature
Operating ............................................................ 10
°
C to 40
°
C (50
°
F to 104
°
F)
Storage ............................................................... -20
°
C to +60
°
C (-4
°
F to +140
°
F)
Humidity ................................................................. 10 % to 90 % non-condensing
Altitude ................................................................... 3,000 m (9,843 ft)
Size (W x H x D) ..................................................... 12.53 cm x 14.86 cm x 4.77 cm (4.94 in x 5.85 in x 1.88 in)
Display ................................................................... LCD Monochrome display
Communication (USB Device Virtual COM Port) .. Mini B connector for service upload of firmware
Power ..................................................................... Lithium-Ion rechargeable, 3.7 V, 10.75 Wh battery, 2900 mAh
Battery Charger ..................................................... 100 V to 240 V, 50/60 Hz input, 6 V/2.5 A output. For best performance, the battery
charger must be connected to a properly grounded ac receptacle.
Battery Life ............................................................ 10 hours (minimum)
Weight .................................................................... 0.29 kg (0.7 lb)
Safety Standards ................................................... EN/IEC 61010-1:2001
Certifications ...........................................................
P, ),
Electromagnetic Compatibility (EMC) ................. EN 61326-1:2006
Detailed Specifications
Oximeter SpO
2
Optical Emitter and Detector
%O
2
O
2
Saturations ..................................................... 80 %, 85 %, 90 %, 95 %, 97 %, 98 %, 99 %, and 100 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range ..........
±
(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy