Put the sensor with the LEDs on the bottom of the
artificial finger. While you put the sensor on the artifact
finger, monitor the signal indicator along the bottom of the
Product display. Adjust the sensor on the finger for
maximum signal strength.

Note

Cable position changes with sensor
manufacturer

How to Set Test Parameters

When you turn on the Product, all parameters are set to
their default values. To change a parameter value, push

or  to move the highlight to a parameter that

you need to change. Push  to move the highlight to
the value of the parameter. Push  or  to
scroll through the values. With the correct parameter
value shown in the display, push  to set the
parameter. Table 5 is a list of all parameters and their
values.

Table 5. Test Parameters

Parameter Values*

SpO2

80%, 85%, 90%, 95%, 97%,
98%, 99%, 100%

HR (Heart Rate)

30, 60, 80, 100, 120, 150, 180,
and 240 BPM

PA (Pulse Amplitude) 0.2%, 2.0%, and 10%

Transmission LG

(Large),

Med (Medium),

and Sm (Small) finger

Artifact

None, Respiration: 2.5%,
Ambient light: 50 or 60 Hz

Type Nonin,

Masimo,

Nellcor, Nihon

Kohden, Mindray, GE-Ohmeda,
Philips/HP, and BCI

Test

Manual, Custom 1, Custom 2,
and Custom 3

* Default values are in bold type.