How to set test parameters – Fluke Biomedical SPOT Light User Manual
Page 20

SPOT Light
Users Manual
8
Put the sensor with the LEDs on the bottom of the
artificial finger. While you put the sensor on the artifact
finger, monitor the signal indicator along the bottom of the
Product display. Adjust the sensor on the finger for
maximum signal strength.
Note
Cable position changes with sensor
manufacturer
How to Set Test Parameters
When you turn on the Product, all parameters are set to
their default values. To change a parameter value, push
or to move the highlight to a parameter that
you need to change. Push to move the highlight to
the value of the parameter. Push or to
scroll through the values. With the correct parameter
value shown in the display, push to set the
parameter. Table 5 is a list of all parameters and their
values.
Table 5. Test Parameters
Parameter Values*
SpO2
80%, 85%, 90%, 95%, 97%,
98%, 99%, 100%
HR (Heart Rate)
30, 60, 80, 100, 120, 150, 180,
and 240 BPM
PA (Pulse Amplitude) 0.2%, 2.0%, and 10%
Transmission LG
(Large),
Med (Medium),
and Sm (Small) finger
Artifact
None, Respiration: 2.5%,
Ambient light: 50 or 60 Hz
Type Nonin,
Masimo,
Nellcor, Nihon
Kohden, Mindray, GE-Ohmeda,
Philips/HP, and BCI
Test
Manual, Custom 1, Custom 2,
and Custom 3
* Default values are in bold type.