How to set custom tests – Fluke Biomedical SPOT Light User Manual
Page 21

SpO2 Functional Tester
How to Set Custom Tests
9
As one example, to change the SpO2 value to 98 %:
1. Push
or to move the highlight to
SpO2: in the display.
2. Push
. The highlight moves to the SpO2
parameter value as shown in Figure 5.
gvh005.jpg
Figure 5. SpO2 Parameter Change
3. Push
or to scroll through the SpO2
parameter values until 98% shows in the display.
Note
As each parameter is set, the artificial finger
outputs the new parameter value immediately.
4. Push
. The highlight moves back to SpO2:
and the value stays set at 98 %.
How to Set Custom Tests
When the Test parameter is set to Manual, no parameter
values are stored. A maximum of three custom tests can
be stored in the Product.
To set up a custom test:
1. Push
or to move the highlight to Test:
in the display.
2. Push
. The highlight moves to the Test
parameter.
3. Push
or to scroll through the test
values. Stop when Custom 1, Custom 2, or Custom 3
shows in the display.
4. Push
.
When the Test parameter is set to Custom 1, Custom 2,
or Custom 3, each parameter you change becomes a
new value for that custom test.