Pacemaker waveform, Arrhythmia – Fluke Biomedical ProSim 2 User Manual
Page 56

ProSim™ 2/3
Users Manual
44
Lead V6 .............................................................. 80 %
Normal Sinus Rhythm ............................................. 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 Universal ECG
Jacks, color-coded to AHA and IEC Standards.
Amplitude ................................................................. 0.05 mV to 0.45 mV (0.05 mV steps), 0.5 mV to 5.5 mV (0.5 mV steps)
Amplitude Accuracy ................................................
±
2 % of setting Lead II. All other leads
±
5 %
ECG Rate .................................................................. 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM
Rate Accuracy ..........................................................
±
1 % of setting
ECG Waveform Selection ........................................ Adult (80 ms) or pediatric (40 ms) QRS duration
Artifact (Superimposed) .......................................... 50 and 60 Hz, muscle, baseline wander, respiration
ST-Segment Elevation ............................................. Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps) Additional steps: +0.05 mV and -0.05 mV
Power-On Default ..................................................... 80 BPM, 1.0 mV, adult QRS, ST-segment elevation of 0 mV, and a P-R interval of 0.16 seconds
Pacemaker Waveform
Pacer-Pulse Amplitude ............................................ 0 (off), 1, 2, 5, 10 mV
±
10 % for lead II (reference lead) with other leads proportional as for performance
waves.
Pacer-Pulse Width ................................................... 0.1, 0.5, 1.0, 1.5, 2.0 ms
±
5 %
Pacing Rate .............................................................. 75 BPM
Paced Arrhythmias .................................................. Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction
Power-On Default ..................................................... Off
Arrhythmia
Baseline NSR ........................................................... 80 BPM
PVC Focus ................................................................ Left focus, standard timing (except where specified)
Supraventricular Arrhythmia .................................. Atrial fibrillation (coarse or fine), atrial flutter, sinus arrhythmia, missed beat (one time), atrial tachycardia,
paroxysmal atrial tachycardia, nodal rhythm, and supraventricular tachycardia
Premature Arrhythmia ............................................. (All one-time events) Premature atrial contraction (PAC), premature nodal contraction (PNC), PVC1 left
ventricular, PVC1 left ventricular – early, PVC1 left ventricular – R on T, PVC2 right ventricular, PVC2
right ventricular – early, PVC2 right ventricular – R on T, and multifocal PVCs