Ecg tests, How to set a performance wave output, R-wave detection test – Fluke Biomedical ProSim 2 User Manual
Page 26

ProSim™ 2/3 
Users Manual 
14
To stop an arrhythmia simulation:
1. Highlight one of the group icons.
2. Push
.
3. Push
or  until the type value shows Off.
ECG Tests
The Product can source pulse, square, triangle, and sine 
waveforms that can be used to verify patient monitors and 
other ECG equipment. These waveforms are used in tests 
for frequency response, sensitivity, gain drift, internal 
calibration, stylus damping, paper speed, linearity, sweep 
speed, and more. 
The Product also sources an R-wave that is used to verify 
that ECG equipment can detect the R-wave part of an 
ECG signal. 
Note
When the Product is set to source a performance 
waveform, respiration and blood pressure 
simulations are disabled. 
How to Set a Performance Wave Output
To set a performance wave on the ECG terminals:
1. In
the
Home screen, push the navigation buttons to
highlight
PERF
.
2. Push
 to show the Performance Wave screen in
the display.
3. Use the parameter selection procedure described in
the Operation section to set the frequency, shape, and 
amplitude of the performance waveform. 
See the detailed specifications for the range of parameter 
values. 
R-Wave Detection Test
You can set the Product to source a normal heart ECG 
signal and vary the amplitude and width of the R-wave 
portion of the waveform. To set the R-wave portion of an 
ECG waveform: 
1. In
the
Home screen, push the navigation buttons to
highlight
PERF
.
2. Push
 to show the Performance Wave screen in
the display.
3. Push the navigation buttons to highlight
RWDET
.
4. Push
 to show the R-Wave Detection screen in
the display.
5. Use the parameter selection procedure described in
the Operation section to set the beats per minute 
(bpm), width, and amplitude of the R-wave. 