Cardiac functions, Ecg functions – Fluke Biomedical ProSim 2 User Manual
Page 22

ProSim™ 2/3
Users Manual
10
2. Push
. The cardiac output screen in Figure 7
shows in the display.
hal003.eps
Figure 7. Cardiac Output Screen
3. To set the injection temperature, push to move the
highlight to the INJ Temp value.
4. Push
or to change the value for the highlighted
parameter. The adjusted simulated value changes the
output signal immediately.
All parameter values in the Product are set with this
procedure. For the parameters that cannot be set, the
highlight will not move to the value of that parameter.
There are two procedures to move between Product
functions. When the home screen is not shown in the
display, Prev, Home, and Next show along the bottom of
the display. To move to the home screen, move the
highlight to Home and push . The home screen
shows all the Product functions. See Figure 5.
An alternative to the Home screen is to use the Prev and
Next selections. The software lets you move through the
Product functions sequentially. The function sequence is:
ECG, Respiration, Blood Pressure, Arrhythmias,
Performance Wave, Pacemaker Wave, Temperature,
Cardiac Output, Fetal Maternal, and Setup.
For example, look at the Cardiac Output screen in
Figure 7. When you move the highlight to Prev and push
, the display shows the Temperature screen. When
you highlight Next and push in the Cardiac Output
screen, the display shows the Fetal/Maternal screen.
Cardiac Functions
The cardiac functions of the Product are ECG, Arrhythmia,
Blood Pressure, Pacemaker, Cardiac Output, and
Performance Waves.
ECG Functions
The ECG function of the Product lets you set five
parameters of an ECG waveform: Rate, Amplitude, Patient
Type, ST, and Artifact. Figure 9 shows a typical setup for
an ECG test on a patient monitor.
To set an ECG parameter:
1. In
the
Home screen, push the navigation buttons to
move the highlight to
ECG
.
2. Push
to show the ECG screen in Figure 8 in the
display.