To set a periodic FHR with IUP:
1. If the Fetal Maternal screen shows in the display, go to
step 3. If not, go to the Home screen of the Product 
and push the navigation buttons to highlight 
FE/MA
.
2. Push
 to show the Fetal Maternal screen.
3. Use the parameter selection procedure to set the FHR,
IUP, and Period parameters
4. Push the navigation buttons to highlight
START
.
5. Push
 to start the test. If the Period parameter is
set to Manual, the simulation stops automatically after 
the IUP wave stops. Each time you push  
another IUP wave starts. If not set to Manual, the IUP 
wave repeats at the frequency set in the Period 
parameter until the simulation is stopped. 
To stop the simulation, highlight
STOP
and push .