Fluke Biomedical ProSim 2 User Manual
Page 34

ProSim™ 2/3
Users Manual
22
To set a periodic FHR with IUP:
1. If the Fetal Maternal screen shows in the display, go to
step 3. If not, go to the Home screen of the Product
and push the navigation buttons to highlight
FE/MA
.
2. Push
to show the Fetal Maternal screen.
3. Use the parameter selection procedure to set the FHR,
IUP, and Period parameters
4. Push the navigation buttons to highlight
START
.
5. Push
to start the test. If the Period parameter is
set to Manual, the simulation stops automatically after
the IUP wave stops. Each time you push
another IUP wave starts. If not set to Manual, the IUP
wave repeats at the frequency set in the Period
parameter until the simulation is stopped.
To stop the simulation, highlight
STOP
and push .