Cardiac output, Cardiac-output test set up – Fluke Biomedical ProSim 2 User Manual
Page 29

Vital Signs Simulator
Cardiac Functions
17
How to Add a Respiration Artifact to the BP 
Waveform 
When the dynamic parameter for a blood pressure channel 
is set to a value other than off, the Product will let you 
move the highlight to the artifact parameter. With the 
highlight on the artifact value, push  or  to toggle the 
value between on and off. 
Each BP channel has a different range of pressure change 
due to the respiration artifact. 
Cardiac Output
The Cardiac Output function electronically simulates the 
dynamic temperature changes of blood that is cooled by 
an injectate. 
Note
Cardiac output measurement devices that use the 
Fick dye injection, Doppler ultrasonography, and 
bioimpedance are not addressed or intended for 
this Product. 
Cardiac-Output Test Set Up
To simulate cardiac output with the Product, a CI-3 
adapter is necessary to connect the monitor to the 
Product. The adapter is shown in Figure 11. Note that the 
injectate temperature thermistor has to be cut off at the 
EUT cable to install the general-purpose connector. 
This module has connections for the cardiac output 
measurement under test and simulates the injectate 
temperature (IT) thermistors at 0 
°
C or 24
°
C. Of the two
connectors on the CI-3 module/cable, the smaller 3-pin
connector is for catheter blood temperature (BT) and is 
standard on most monitors. 
Note
This 3-pin catheter BT connector is compatible 
with the standard Baxter (Edwards) BT catheter 
and equivalent catheters available from other 
manufacturers such as Viggo-SpectraMed and 
Abbott (Sorenson). 
The larger 4-pin connector supplies the simulated injectate 
temperature. The 10-turn 100 k
Ω
potentiometer enables
adjustment of the injectate temperature to 0
°
C or 24
°
C.
The 4-pin IT thermistors connector is not standard on all 
monitors. A general function connector that you can 
connect to the device under test (DUT) injectate cable is 
also available. 
Note
A DUT cardiac output cable changed for this test 
must not be used in clinical applications. 