Other functions, Respiration functions, Temperature – Fluke Biomedical ProSim 2 User Manual
Page 36

ProSim™ 2/3 
Users Manual 
24
Other Functions
The Product also simulates respiration and temperature. 
This section contains the procedures to set up the Product 
for these two functions. 
Respiration Functions
The Respiration function lets you set five parameters of 
the respiratory waveform: Rate, Impedance, Baseline 
Impedance, Lead selection (left arm or left leg), and 
Apnea. To set the respiration waveform: 
1. In
the
Home screen, push the navigation buttons to
move the highlight to
RESP
.
2. Push
to show the respiration screen in the
display.
See the Function Navigation and Parameter Selection 
section to learn how to set the respiration parameter 
values. See the detailed specifications for range of 
parameter values. These parameters change the 
waveform that appears at the ECG posts along the top of 
the Product. 
When the Apnea parameter is set to 12, 22, or 32, the 
apnea event starts immediately. When the event ends, the 
parameter is set to Off. You must set the parameter to 12, 
22, or 32 to start another apnea event. 
The values set for the baseline and lead parameters when 
the Product is turned off, become the power-up default 
values. 
Temperature
Temperatures simulated by the Product are compatible 
with Yellow Springs, Inc. (YSI) Series 400 and 700 probes. 
The type of cable connected to the temperature jack sets 
the type of temperature probe simulated. Connect the 
temperature input of the UUT to the Temperature jack as 
shown in Figure 14. 
To set temperature:
1. In
the
Home screen, push the navigation buttons to
move the highlight to
TEMP
.
2. Push
to show the temperature screen in the
display.
See the Function Navigation and Parameter Selection 
section to learn how to set the temperature parameter 
value. See the detailed specifications for range of 
parameter values. These parameters change the 
temperature signal at the temperature connector.