Performance specifications – Fluke Biomedical INCU Incubator Analyzer User Manual
Page 60

INCU
Operators Manual
A-2
Performance Specifications
This appendix provides the device specifications for the INCU Incubator
Analyzer. For more information, please contact your Fluke Biomedical
Service representative.
Temperature Measurements
Conduction:
1 sensor in contact with mattress
(bottom of unit)
Convection
(Inside the incubator):
3 sensors (T1, T3, T4)
Radiation:
1 sensor (T2) connected to the so-called “radiant
baby”
Convection
(Inside or outside the
incubator):
1 sensor T2
Measuring Range:
5
°C to 70°C
Resolution:
0.1
°C
Accuracy:
Reading Value
±0.5°C ± 1 LSB for the range
from 25
°C to 40°C
Storage Temperature
-20°C to 50°C
Operating Temperature:
10°C to 40°C
Dimensions
:
27 x 14 x 20 cm
Relative Humidity
Measuring Range:
0% RH to100% RH
Resolution:
0.1% RH
Accuracy:
±5% RH (for range 0 to 90% RH
@ 25
°C to 40°C) or,
± 5.3 % RH (for range 0 to 100% RH
@ 25
°C to 40°C)
Humidity must be without condensation.
Air Flow
Measuring Range:
0.1-0.7 m/s
Resolution: 0.01
m/s
Accuracy:
From 0.1 m/s to 0.5 m/s reading
±0.1 m/s @ temperature 25°C to 40°C
and humidity 50% RH
± 15% RH