Incu operators manual, Figure 3-4. placement of temperature sensors t1-t3, T1, t2, and t3 convection sensors – Fluke Biomedical INCU Incubator Analyzer User Manual
Page 32: Humidity sensor (fixed)

INCU
Operators Manual
3-6
T1, T2, and T3 Convection Sensors:
Figure 3-4 shows the recommended
placement for temperature sensors T1-T3 for closed incubators utilizing
forced convection. With the sensors mounted to the provided clips,
temperature accuracy, response, and uniformity can be monitored throughout
the area occupied by the infant at the height of approximately 10 cm above
the mattress as specified by standards. In this case, T2 is being utilized as a
convection temperature sensor.
Humidity Sensor (Fixed):
The humidity sensor is fixed and is located on the
inside of the right-hand cover as shown in Figure 3-4.
INCU
INCUBATOR A
NALYZER
DO NOT USE
WITH OXYG
EN
NE PAS UTIL
ISER EN PR
ESENCE
D'OXYGENE
T3
AIR FLOW
RELATIVE
HUMIDITY
CONVECTIO
N
T1
T2
CONVECTION
RADIANT/
CONVECTIO
N
!
EVENT
SELECT
T3 Convection
Humidity
Air Flow
T1 Convection
On/Off Switch
Temperature
Probe Holder
T2 Convection
or Conduction
Figure 3-4. Placement of temperature sensors T1-T3
Note:
To achieve a better temperature slew rate, for T1, T2, it
is preferable to clip on the sensor cable instead of the
body sensor.