Transferring data to the pc software – Fluke Biomedical INCU Incubator Analyzer User Manual
Page 43

Incubator Analyzer
Transferring Data to the PC Software
4-5
6. Press the
Events
key for approximately
1 second to mark a particular value or a pause
in the recording sequence. As a result, the
action is recorded and then displayed on the
History Graph Sheet and Parameter Numerical
List (see Figures B-2 and B-3 in Appendix B,
Report Examples).
7. Describe these events on the
monitoring
sheet
if one was printed before data
acquisition. It is possible to record up to five
events.
When the acquisition phase is complete,
end of
measurements
will be displayed.
Transferring Data to the PC Software
After data acquisition is complete, data transfer can be executed. Data
transfer is the act of transferring the data in the INCU from the previous
acquisition to a PC running INCU PC software. When data is transferred to a
PC, it can be displayed in graphical or numerical form, and/or saved as a file
for future reference (see Appendix B for examples of graphs and numerical
data). Numerical data can be entered into Microsoft® Excel or some other
spreadsheet program to customize your own graphs.
To initiate a data transfer from the INCU, connect the INCU to a PC running
the INCU PC software via the RS-232 port utilizing the supplied RS-232
cable. Click on connect and transfer as shown in Figure 4-4 to initiate data
transfer. During the transfer phase, the mouse pointer indicates the progress
of data transfer.
Figure 4-4. INCU file transfer menu